FDA recall Z-2282-2023

Covidien, LLC · Class II · device

Product

Covidien Cytosponge Cell Collection Device, REF: CYTO-201; and Covidien Cytosponge Cell Collection Kit, REF: CYTO-KIT-USM

Reason for recall

Cell Collection Device may be at increased risk of the sponge detaching from the string during removal of the device from the patient, which could lead to device fragments in patient, obstruction, airway obstruction, secondary intervention, secondary intervention (with the primary procedure), supraglottic airway obstruction, and aspiration.

Distribution

Worldwide - US Nationwide distribution in the states of NY, MN, VA, WI and the countries of Canada, Netherlands, Poland, United Kingdom.

Key facts

Status
Ongoing
Initiation date
2023-06-02
Report date
2023-08-09
Voluntary/Mandated
Voluntary: Firm initiated
Location
Santa Clara, CA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2282-2023