# FDA recall Z-2288-2018

> **Cenorin, LLC** · Class II · device recall initiated 2018-04-09.

## Product

HLD Systems 600 Series Washer/Pasteurizer, UDI (01) 0085249007009 (11) 160929 (21) 61160; Model 610HT, 115V & Model 610HT, 230V    Product Usage:  Clean and thermally disinfect medical devices using full immersion pasteurization

## Reason for recall

There is a supplier quality issue with the terminal blocks which may cause the terminal block to overheat, char and/or burn.

## Distribution

Worldwide Distribution - US Nationwide in the states of CA, FL, IN, NC, NE, UT & WV; and countries of  Malaysia and Taiwan

## Key facts

- **Recall number:** Z-2288-2018
- **Recalling firm:** Cenorin, LLC
- **Classification:** Class II
- **Product type:** device
- **Status:** Terminated
- **Initiation date:** 2018-04-09
- **Report date:** 2018-07-04
- **Termination date:** 2019-09-13

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Kent, WA, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2288-2018

## Citation

> AI Analytics. FDA recall Z-2288-2018. Retrieved 2026-06-09 from https://api.ai-analytics.org/recall/Z-2288-2018. Source: US FDA. Licensed CC0.

---

*[Dataset catalog](/datasets/) · [FDA recalls dataset](/datasets/fda-recalls) · CC0 1.0*