Trilogy Evo O2, Product numbers DE2100X13B, DS2100X11B, EE2100X15B, ES2100X15B, EU2100X15B, FR2100X14B, FX2100X15B, IA2100X15B, IN2100X15B, IN2100X19, IT2100X21B, JP2100X16B, LA2100X15B, ND2100X15B, RDE2100X13B, SP2100X26B
Two software issues have been identified related to pressure increase in the device: The first issue is described as "Infant/Pediatric EFS Calibration Pressure Increase" and the second issue is described as "Pressure Drift (Continuous Usage).
Worldwide distribution - US nationwide distribution.
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2288-2021