FDA recall Z-2289-2018

Zimmer Biomet, Inc. · Class II · device

Product

Vanguard Complete Knee System; E1 Vanguard PS Tibial Bearing; Model Numbers EP-183664, EP-183642 Product Usage: The product is intended for use in knee joint replacement arthroplasties.

Reason for recall

Two lots of tibial bearings were commingled. There is a possibility that a 14mm tibial bearing is packaged in a box labeled as 12 mm tibial bearing and vice versa. This may lead to extension of surgery to find another implant.

Distribution

Distributed to accounts in CA, CT, GA, IA, IN, LA, MA, MD, MN, MS, NJ, OH, PA, TN, TX, UT, VA, WI. Foreign distribution to Taiwan, Japan, and the Netherlands.

Key facts

Status
Terminated
Initiation date
2018-05-24
Report date
2018-07-04
Termination date
2020-04-07
Voluntary/Mandated
Voluntary: Firm initiated
Location
Warsaw, IN, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2289-2018