# FDA recall Z-2295-2025

> **Merit Medical Systems, Inc.** · Class II · device recall initiated 2025-06-05.

## Product

Custom Manifold Kit:  REF: K09-02239C, K09-02249UP, K09-09110B, K09-10294A, K09-11227A, K09-11456B, K09-12123B, K09-13342AP, K09-13391BP, K09-13548, K09-14000A, K09T-12020D, K09T-12275B

## Reason for recall

High pressure tubing may relax over time causing it to partially separate from the connector resulting in a void that can trap air in the line. Use of the affected product may result in the trapped air being introduced into the patient.

## Distribution

US Nationwide distribution in the states of OH, CA, SD, VA, AZ, MT, TX, MI, FL, TN, PA, NJ, MN, ND, IA, LA, UT, KY, IL, OK, ME, IN, NY, NM, SC, AL, ID, CT, WI, GA, MA, NC, AR, WA.

## Key facts

- **Recall number:** Z-2295-2025
- **Recalling firm:** Merit Medical Systems, Inc.
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2025-06-05
- **Report date:** 2025-08-20

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** South Jordan, UT, United States