FDA recall Z-2302-2020

Siemens Medical Solutions USA, Inc · Class II · device

Product

Multi Display Manager in the Artis Systems (Image-Intensified Fluoroscopic X-Ray System) as follows: Artis Q floor Model: 10848280; Artis Q ceiling Model: 10848281; Artis Q biplane Model:10848282; Artis zee ceiling Model: 10094137; Artis zee floor Model: 10094135 Product Usage: Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.

Reason for recall

Artis systems containing a Multi Display Manager of a specific lot, the power distributor may lead to an increased electrical contact resistance or the interruption of electrical contact, and result in planned procedures to be terminated and performed on an alternative x-ray system

Distribution

US Nationwide distribution.

Key facts

Status
Terminated
Initiation date
2020-04-30
Report date
2020-06-17
Termination date
2020-09-11
Voluntary/Mandated
Voluntary: Firm initiated
Location
Malvern, PA, United States

Primary source