FDA recall Z-2308-2021

Hitachi Healthcare Americas Corporation · Class II · device

Product

Model L43K Intraoperative Probe-For the diagnostic ultrasound evaluation during robotic/non-robotic intra-operative and laparoscopic procedures

Reason for recall

The screw cover at the tip of the probe may fall off after a cleaning and sterilization process. Users advised to conduct a preventive inspection of the probe before and after sterilization/disinfection/cleaning and before and after an interoperative procedure.

Distribution

Worldwide distribution - US Nationwide and the country of Canada.

Key facts

Status
Ongoing
Initiation date
2021-06-30
Report date
2021-08-25
Voluntary/Mandated
Voluntary: Firm initiated
Location
Twinsburg, OH, United States

Primary source