FDA recall Z-2315-2024

Baxter Healthcare Corporation · Class II · device

Product

iLED 7 surgical lights, Model Numbers: a) 4068110 (Ceiling Single Surgical Light); b) 4068120 (Mobile Surgical Light); c) 4068140 (Pendant Surgical Light); d) 4068210 (Ceiling Duo Surgical Lights); e) 4068310 (Ceiling Trio Surgical Lights); f) 4068410 (Ceiling Quad Surgical Lights)

Reason for recall

Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction due to customer reports of thermal skin injuries when using the iLED 7 surgical light system products listed below. Specific VidiaPort and Trulight product codes are also impacted as they are equipped with iLED 7 light-heads

Distribution

Worldwide distribution.

Key facts

Status
Ongoing
Initiation date
2024-06-04
Report date
2024-07-17
Voluntary/Mandated
Voluntary: Firm initiated
Location
Deerfield, IL, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2315-2024