FDA recall Z-2322-2017

Philips Electronics North America Corporation · Class II · device

Product

Philips DuraDiagnost X-ray systems: DuraDiagnost (R1.0, Product Number 712210) DuraDiagnost (R2.0, Product Number 712211) DuraDiagnost (R3.0, Product Number 712211)

Reason for recall

Tube arm assembly could fall down due to cracked welding joints

Distribution

Worldwide Distribution - US Distribution to the states of : Ca and Oh., and to the countries of: Australia, Libya, China, Mayotte, Mexico, France, United Kingdom, Germany, Ecuador, Australia, Spain, Switzerland, Macedonia, Uzbekistan, Kuwait, Malaysia, Netherlands, Argentina, Chile, Colombia, Egypt, United Arab Emirates, Saudi Arabia, Hungary, Peru, Poland, Indonesia, Burkina Faso, Kazakhstan, Turkey and Slovenia.

Key facts

Status
Terminated
Initiation date
2017-05-31
Report date
2017-06-21
Termination date
2018-08-24
Voluntary/Mandated
Voluntary: Firm initiated
Location
Andover, MA, United States