FDA recall Z-2323-2017

GE Healthcare, LLC · Class II · device

Product

Infant Warmer System (IWS)

Reason for recall

GE Healthcare has recently become aware of a potential safety issue related to loose screws in the Heater Head of the Infant Warmer System (IWS). Hot screws from the Heater Head of the IWS could fall onto the bed if the Heater Head assembly has been improperly serviced. This situation can be clinically hazardous because thermal injury to a patient could result. Two injuries have been reported as a result of this issue.

Distribution

Worldwide Distribution - USA (nationwide) Distribution.

Key facts

Status
Terminated
Initiation date
2017-05-05
Report date
2017-06-21
Termination date
2018-07-12
Voluntary/Mandated
Voluntary: Firm initiated
Location
Waukesha, WI, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2323-2017