FDA recall Z-2323-2024

Baxter Healthcare Corporation · Class II · device

Product

Welch Allyn CP150 Electrocardiograph with Spirometry Option, Product Code: CP150A-1ENB. The Electrocardiograph is one of the tools that clinicians use to evaluate, diagnose, and monitor patient cardiac function.

Reason for recall

During manufacturing, the step to install an Electromagnetic Interference (EMI) Absorber was omitted. The purpose of the EMI absorber is to prevent EMI leakage that might potentially affect other devices in the vicinity or the device itself may be vulnerable to EMI resulting in device malfunction or unavailability.

Distribution

Distribution to Illinois.

Key facts

Status
Ongoing
Initiation date
2024-06-03
Report date
2024-07-17
Voluntary/Mandated
Voluntary: Firm initiated
Location
Deerfield, IL, United States