FDA recall Z-2324-2018

St Jude Medical Inc. · Class II · device

Product

Quadra Assura, Sterile EO, Model #/ Part #: CD3265-40/100042335, 100059846; CD3265-40Q/100042336, 100059847; CD3267-40/100042956, 100043569, 100059961, 100078810; CD3267-40Q/100043132, 100043539, 100059917, 100059956, 100078904; CD3365-40C/100080435, 100080436; CD3365-40Q/100080426, 100080427; CD3367-40/100078813, 100079497, 100096097; CD3367-40C/100078597, 100079174, 100079450, 100079490, 100105524; CD3367-40Q/100078851, 100079212, 100079504, 100096098; CD3367-40QC/100078656, 100079193, 100079498, 100079524, 100105513, 100105514, 100138185, 100138213; CD3385-40C/100080778; CD3385-40QC/100080780, 100105489

Reason for recall

The firm is releasing new firmware for high voltage implantable cardiac devices (ICD) and cardiac resynchronization therapy defibrillators (CRT-D) intended to strengthen security by providing an additional layer of protection against unauthorized device access. For older generation devices not capable of accepting the firmware, the firm is advising customers of the option to permanently disable the RF communication capability in these devices.

Distribution

U.S. Nationwide distribution

Key facts

Status
Terminated
Initiation date
2018-04-16
Report date
2018-07-04
Termination date
2024-03-08
Voluntary/Mandated
Voluntary: Firm initiated
Location
Sylmar, CA, United States