ENDOTAK RELIANCE G/SG, transvenous defibrillation lead, implantable cardioverter defibrillator (non-CRT); Model Nos. 0160, 0161, 0162, 0163, 0164, 0165, 0166, 0167, 0170, 0171, o182, 0173, 0174, 0175, 0176, 0177, 0180, 0182, 0183, 0184, 0185, 0186, 0187
Affected lead models may experience gradually rising LVSI measurements in association with calcification of the defibrillation lead coils. The issue may lead to early lead replacement. Most serious harm is death or need for cardiac resuscitation.
Worldwide Distribution. US nationwide. International distribution worldwide.
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2325-2025