FDA recall Z-2327-2025

Boston Scientific Corporation · Class I · device

Product

RELIANCE 4-FRONT, endocardial cardioversion/defibrillation and pace/sense lead, Permanent Defibrillator Electrodes; Model Nos. 0654, 0655, 0657, 0658, 0682, 0683, 0685, 0686, 0692, 0693, 0695, 0696

Reason for recall

Affected lead models may experience gradually rising LVSI measurements in association with calcification of the defibrillation lead coils. The issue may lead to early lead replacement. Most serious harm is death or need for cardiac resuscitation.

Distribution

Worldwide Distribution. US nationwide. International distribution worldwide.

Key facts

Status
Ongoing
Initiation date
2025-07-24
Report date
2025-08-27
Voluntary/Mandated
Voluntary: Firm initiated
Location
Saint Paul, MN, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2327-2025