# FDA recall Z-2332-2024

> **Barco N.V.** · Class II · device recall initiated 2024-04-24.

## Product

MNA-420 ENC HDMI, Material Number K9303311; cathode ray tube display

## Reason for recall

We have identified a potential issue with the SFP+ module supplied by Molex (Part Number: B620050) when connected to the ETH0 slot of either the MNA-6x0 ENC HDMI or MNA-4x0 ENC HDMI. In some instances, this connection may render the first HDMI channel inaccessible, making both the channel and the respective MNA ENC unit unusable. Additionally, readings from the module become unavailable when this issue arises.    Similarly, when this Molex module is connected to the ETH0 slot of an MNA-620 ENC ANA, readings may also become inaccessible.

## Distribution

US:  FL

## Key facts

- **Recall number:** Z-2332-2024
- **Recalling firm:** Barco N.V.
- **Classification:** Class II
- **Product type:** device
- **Status:** Completed
- **Initiation date:** 2024-04-24
- **Report date:** 2024-07-17

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Kortrijk, N/A, Belgium

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2332-2024

## Citation

> AI Analytics. FDA recall Z-2332-2024. Retrieved 2026-07-26 from https://api.ai-analytics.org/recall/Z-2332-2024. Source: US FDA. Licensed CC0.

---

*[Dataset catalog](/datasets/) · [FDA recalls dataset](/datasets/fda-recalls) · CC0 1.0*