FDA recall Z-2338-2024

Maquet Cardiovascular, LLC · Class II · device

Product

The VASOVIEW HEMOPRO 2 Endoscopic Vessel Harvesting System is designed for use in conjunction with the 7 mm Extended Length Endoscope. The Harvesting Cannula has four lumens which house the Endoscope, C-Ring, distal lens washer tube, and harvesting tool for cutting and cauterizing vessel branches. The harvesting tool has two curved jaws. One of the jaws contains the heating elements for branch cutting and cauterization and spot cautery. An area near the tip of the convex side of the jaw is used for spot cautery. The activation toggle is used to control the jaws to activate the heating elements. Positioning of the device, cutting, and cauterization are performed under endoscopic visualization.

Reason for recall

There were 27 complaints between March 22, 2024, and April 30, 2024, related to VH-3500 and VH-4000, reporting that the C-Ring wire was straightened (due to insufficient bend radius) and that the distance between C-Ring and Harvesting Tool was closer than normal. No adverse events have been reported to date in direct association with this issue.

Distribution

NationwideForeign: AT DE DK ES GB HK IT NL ZA

Key facts

Status
Ongoing
Initiation date
2024-05-31
Report date
2024-07-17
Voluntary/Mandated
Voluntary: Firm initiated
Location
Wayne, NJ, United States