FDA recall Z-2344-2021

Philips North America Llc · Class II · device

Product

Ingenia Elition X-Magnetic Resonance Medical Electrical Systems indicated for use as a diagnostic device. The system can produce cross-sectional images, spectroscopic images and/or spectra in any orientation of the internal structure of the head, body or extremities Model Number: 781358

Reason for recall

Pressure relief valve 42 psi installed in the magnet assembly instead of a 3.75 psi valve, may result in release of helium gas and cause injury

Distribution

Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, MD, MD, NH, NY, OH, TN, TX, VA and the countries of Canada, Mexico, Australia, Austria, Chile, China, Czech Republic, Denmark, France, Germany, India, Indonesia, Italy, Japan, Korea, Republic of Lebanon, Netherlands, Romania, Russian Federation, Saudi Arabia, Spain, Turkey, Ukraine, United Kingdom.

Key facts

Status
Terminated
Initiation date
2021-07-30
Report date
2021-09-01
Termination date
2023-07-25
Voluntary/Mandated
Voluntary: Firm initiated
Location
Cambridge, MA, United States

Primary source