FDA recall Z-2348-2019

Thoratec Corp. · Class II · device

Product

Thoratec HeartMate 3 LVAS Implant Kit, Sterile, REF 106524/106524US, RX only, Left Ventricular Assist System

Reason for recall

Outflow Graft leaking at the pump connection during the implantation process. If the C-Ring is insufficiently seated, it renders the device nonfunctional because the Outflow Graft connection will leak during the implant procedure.

Distribution

Worldwide distribution. US nationwide, Australia, Austria, Brazil, Canada, Cayman Islands, Chile, Colombia, Egypt, France, Germany, Greece, Singapore, Iran, Italy, Kuwait, Lebanon, Romania, Switzerland, Thailand, United Kingdom.

Key facts

Status
Terminated
Initiation date
2019-03-30
Report date
2019-08-28
Termination date
2020-08-24
Voluntary/Mandated
Voluntary: Firm initiated
Location
Pleasanton, CA, United States

Primary source