Randox RX Series Urea Enzymatic Kinetic Assay Ref. Numbers: (1) UR3825 (2) UR3873 (3) UR8334 (4) UR8070 (Modena)
Carryover from the Direct LDL-cholesterol assay on the Rx instruments may result in a negative bias, up to -11%, in quality control and patient samples, in urea assays results if the urea assay is run immediately after the Direct LDL-cholesterol assay. This may lead to a delay in reporting results or in falsely decreased urea test results.
US Nationwide distribution in the states of CA, DC, GA, OH, OK, MN, MO, NJ, VA and Puerto Rico.
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2362-2023