FDA recall Z-2362-2023

Randox Laboratories Ltd. · Class II · device

Product

Randox RX Series Urea Enzymatic Kinetic Assay Ref. Numbers: (1) UR3825 (2) UR3873 (3) UR8334 (4) UR8070 (Modena)

Reason for recall

Carryover from the Direct LDL-cholesterol assay on the Rx instruments may result in a negative bias, up to -11%, in quality control and patient samples, in urea assays results if the urea assay is run immediately after the Direct LDL-cholesterol assay. This may lead to a delay in reporting results or in falsely decreased urea test results.

Distribution

US Nationwide distribution in the states of CA, DC, GA, OH, OK, MN, MO, NJ, VA and Puerto Rico.

Key facts

Status
Ongoing
Initiation date
2023-06-13
Report date
2023-08-16
Voluntary/Mandated
Voluntary: Firm initiated