FDA recall Z-2374-2024

CareFusion 303, Inc. · Class II · device

Product

BD Pyxis: MedStation ES (Main), REF: 323; MedStation ES Tower, REF: 352; Anesthesia Station ES, REF: 327; CII Safe ES, REF: 1116-00; CII Safe ES, Desktop PC, Wired Scanner, REF: 107-254-01; CII Safe ES, Desktop PC, Wireless Scanner, REF: 107-255-01; CII Safe ES, Desktop PC, Wired Scanner 1.7.3, REF: 107-256-01; CII Safe ES, Desktop PC, Wireless Scanner 1.7.3, REF: 107-257-01

Reason for recall

If automated dispensing cabinets have specific software versions, and Component Manager was configured to "installed mode", and software patch KB 5033688 was installed in the production environment, then this will will lead to an error that prevents the user from accessing the dispensing software application, which may delay accessing medication.

Distribution

Worldwide - US Nationwide distribution including in the states of OR, KS, NE, LA, ND, OK, AL, AK, WI, WV, WA, NV, PR, TX, CO, OH, CA, MO, IA, AZ, IN, AR, IL, MN, HI, NM, FL, SD, UT, NY, VA, GA, MA, NC and the countries of TW, CA.

Key facts

Status
Ongoing
Initiation date
2024-05-02
Report date
2024-07-24
Voluntary/Mandated
Voluntary: Firm initiated