FDA recall Z-2389-2018

Roche Diagnostics Corporation · Class II · device

Product

Elecsys Vitamin D total II, Vitamin D Test System, Material number 07028148190 Product Usage: This assay is intended for the quantitative determination of total 25-hydroxyvitamin D in human serum and plasma. This assay is to be used as an aid in the assessment of vitamin D sufficiency in adults. The electrochemiluminescence binding assay is intended for use on the immunoassay analyzer.

Reason for recall

The device may give a falsely elevated result that is non-reproducible. If vitamin D level is monitored by laboratory tests and adapted based on the result, a dosage reduction of vitamin D supplementation cannot be excluded, and as a consequence, vitamin D deficiency may develop over time.

Distribution

US Nationwide distribution.

Key facts

Status
Terminated
Initiation date
2018-04-27
Report date
2018-07-18
Termination date
2020-05-11
Voluntary/Mandated
Voluntary: Firm initiated
Location
Indianapolis, IN, United States

Primary source