FDA recall Z-2406-2021

Elekta Inc · Class II · device

Product

Monaco RTP System, Software Builds 5.40.00, 5.40.01, 5.40.02, and 5.51.10

Reason for recall

If a couch is removed while a plan is loaded, and then added back to the Studyset when the plan is unloaded, it is possible to calculate dose with the couch unassigned to the beams.

Distribution

Distribution in US - IA, NJ, NY, TN, PA, TX, and WI OUS distribution in Australia, Bahrain, Belgium, Canada, China, Denmark, France, Germany, Hong Kong, Italy, Japan, Korea, Netherlands, Spain, Sweden, Switzerland, Thailand, Turkey and United Kingdom

Key facts

Status
Ongoing
Initiation date
2021-07-26
Report date
2021-09-08
Voluntary/Mandated
Voluntary: Firm initiated
Location
Saint Charles, MO, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2406-2021