FDA recall Z-2412-2018

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) · Class II · device

Product

Percepta CRT-P MRI SureScan, REF W1TR01 Product Usage: The CRT-P models are implantable pulse generators with cardiac resynchronization therapy (CRT-P). These devices monitor and regulate the patient s heart rate by providing single or dual chamber rate responsive bradycardia pacing, sequential biventricular pacing, and atrial tachyarrhythmia therapies. These devices are indicated for heart failure patients who may have atrial arrhythmias or AV block.

Reason for recall

Potential for a device reset to occur in Percepta CRT-P MRI SureScan and Percepta Quad CRT-P MRI SureScan due to a timing interaction between the EffectivCRT Diagnostic and the Ventricular Safety Pacing feature (VSP).

Distribution

Worldwide Distribution - US Nationwide

Key facts

Status
Terminated
Initiation date
2018-06-21
Report date
2018-07-18
Termination date
2020-08-19
Voluntary/Mandated
Voluntary: Firm initiated
Location
Mounds View, MN, United States