FDA recall Z-2424-2020

GE Healthcare, LLC · Class II · device

Product

CARESCAPE Patient Data Module (PDM) used with the following monitors: CARESCAPE B450/B650/B850, SOLAR 8000M/I and Transport Pro, Model Numbers: 2042084-001 2094504-001 2042084-001-046050 2042084-001-208621 2042084-001-045915 2042084-001-215164 2042084-001-078474 2042084-001-053034 2042084-001-215165 2042084-001-079462 2042084-001-047389 2042084-001-112198 2042084-001-041673 2042084-001-227238 2042084-001-071564 2042084-001-503824 2042084-001-119839 2042084-001-569269 2042084-001-041674 2042084-001-047549 2042084-001-514651 2042084-001-163572 2042084-001-583721 2042084-001-427592 2042084-001-527358 2042084-001-583723 2042084-001-317166 2042084-001-583724 2042084-001-583715 2042084-001-583718 2042084-001-055156 2042084-001-583722 2042084-001-041835 2042084-001-074124 2042084-001-583706 2042084-001-583707 2042084-001-519926 2042084-001-583717 2042084-001-583713 2042084-001-207863 2042084-001-460604 2042084-001-079192 2042084-001-197663 2042084-0

Reason for recall

The CARESCAPE Patient Data Module (PDM) is used with the following monitors: CARESCAPE B450/B650/B850, SOLAR 8000M/I and Transport Pro. If Pace detection is turned ON, on the monitor, and if an Automated External Defibrillator (AED) is used to perform defibrillation, the low frequency signals deployed by the AED, cause the PDM to issue a false flat line ECG waveform and alarm Asystole to be displayed on the monitor.

Distribution

Worldwide Distribution

Key facts

Status
Ongoing
Initiation date
2020-05-18
Report date
2020-07-01
Voluntary/Mandated
Voluntary: Firm initiated
Location
Waukesha, WI, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2424-2020