# FDA recall Z-2426-2024

> **Radiometer Medical ApS** · Class II · device recall initiated 2024-05-31.

## Product

ABL800 FLEX  Model Numbers 393-800 and 393-801    The ABL800 FLEX analyzers are intended for:   " In Vitro testing of samples of whole blood for the parameters pH, pO2, pCO2, cK+, cNa+, cCa2+, cCl , cGlu, cLac, cCrea, ctBil, and co-oximetry parameters (ctHb, sO2, and the hemoglobin fractions FO2Hb, FCOHb, FMetHb, FHHb and FHbF)   " In vitro testing of samples of expired air for the parameters pO2 and pCO2   " In vitro testing of pleura samples for the pH parameter.

## Reason for recall

Issue with analyzer when the pH of the calibration solution decreases during the in-use period potentially because of bacterial growth in the calibration solution bottles.  This may result in a probability of reporting biased out-of-specification pH results on blood samples.

## Distribution

Worldwide - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The countries of Canada, Afghanistan, Albania, Algeria, Argentina, Bahrain, Bangladesh, Belarus, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Chile, Colombia, Cuba, Cyprus, Dominican Republic,  Ecuador, Egypt, Eritrea, Ethiopia, Greece, Guatemala, Honduras, Iraq, Iran, Israel, Jamaica, Jordan, Kenya, Kuwait, Lebanon, Libya, Lithuania, Malta, Mauritius, Moldova, Nepal, Nicaragua, Nigeria, Oman, Pakistan, Paraguay, Peru, Philippines, Puerto Rico, Qatar, Romania, Saudi Arabia, Serbia, Slovenia, Sri Lanka, Suriname, Syria, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Uganda, Ukraine, United Arab Emirates, Uruguay, Uzbekistan, Venezuela, Vietnam, Yemen, Zimbabwe Australia,  New Zealand, New Caledonia, Tokelau, Austria, Belgium, Netherland

## Key facts

- **Recall number:** Z-2426-2024
- **Recalling firm:** Radiometer Medical ApS
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2024-05-31
- **Report date:** 2024-07-31

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Bronshoj, Denmark