# FDA recall Z-2433-2018

> **AESDEX** · Class II · device recall initiated 2018-04-02.

## Product

AesDex, LLC  Cardica C-Port xA PLUS Anastomosis System. Catalog Number FG-000100.      The Cardica C-Port xA PLUS Anastomosis System, delivers a series of clips that create an anastomosis between a small target vessel (e.g. coronary artery) and conduit (e.g. saphenous vein graft). The stainless steel clips create a complete end-to-side anastomosis that is functionally equivalent to a hand-sutured interrupted stitch anastomosis. The system consists of one Anastomosis Device and one Retractor Mount.

## Reason for recall

Following pressurization of the devices, damage to the device handle mechanism may occur, accompanied by a loud noise and potential dispersion of plastic fragments and components.

## Distribution

U.S. and international: Belgium, Germany, Italy and Netherlands. No gov accounts.

## Key facts

- **Recall number:** Z-2433-2018
- **Recalling firm:** AESDEX
- **Classification:** Class II
- **Product type:** device
- **Status:** Terminated
- **Initiation date:** 2018-04-02
- **Report date:** 2018-07-25
- **Termination date:** 2021-04-27

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Redwood City, CA, United States