FDA recall Z-2435-2019

Integra LifeSciences Corp. · Class II · device

Product

Integra Jarit GEMINI CLAMP 9 1/4 inch - used in vascular and cardiothoracic procedures used for clamping and occluding vessels. Cat. #140-332

Reason for recall

A small crack at the box lock corner may after multiple cleaning and sterilization cycles produce an increased risk of corrosion, weakening the instrument and resulting in premature need for replacement and/or breakage of the instrument

Distribution

US Nationwide distribution.

Key facts

Status
Terminated
Initiation date
2019-08-07
Report date
2019-09-11
Termination date
2020-06-24
Voluntary/Mandated
Voluntary: Firm initiated
Location
Plainsboro, NJ, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2435-2019