FDA recall Z-2438-2018

Siemens Healthcare Diagnostics, Inc. · Class II · device

Product

Dimension Vista¿ MULTI 2 SDIL, KD694, SMN# 10483586

Reason for recall

The diluent may have an incomplete slit on the septum in the cap of the vial. A diluent cap with a defective septum has the potential to cause erroneously elevated results if a sample is autodiluted onboard the system, due to a reduced volume of diluent being pipetted from the vial.

Distribution

The products were distributed to the following US states: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The products were distributed to Guam, Saipan, and Virgin Islands. The products were distributed to the following countries: Australia, Austria, Bahamas, Belgium, Brazil, Canada, Canary Islands, Croatia, Czech Republic, Denmark, France, Germany, Italy, Japan, Lebanon, Netherlands, Poland, Portugal, Qatar, Russian Federation, Slovakia, Slovenia, Saudi Arabia, South Korea, Spain, Switzerland, and United Arab Emirates.