FDA recall Z-2440-2021

Copan Italia · Class II · device

Product

COPAN FLOQSwabs, STERILE.EO, for drawing clinical samples. CE0123, for the following reference numbers REF:501CS01, Minitip, sterile Single Wrapped, Molded bp 80mm; REF: 20226, P/N 501CS01Q100, Minitip, Sterile, Single Wrapped, bp 60MM, BUNDLES OF 20 pcs; REF: 501CS01.Q20, P/N 1127700, Minitip Applicator Sterile Ind. wrapped., bp 60mm, bundles of 20; REF: 502CS01, Regular, Sterile single Wrapped, Molded bp 80mm; REF: 503CS01, Flexible, Sterile Single Wrapped, Molded bp 100mm; REF: 220252, P/N 503CS01.BD, Flexible, Sterile Single Wrapped Molded bp 100 mm; REF: 503CS01/503CS01.BX, Flexible, Sterile Single Wrapped, Molded bp 100mm REF: 50U004DS, hDNA Free, Regular sterile in dry tube Molded Break Point 20 mm; REF: 50U009DS02, hDNA Free, Regular, Sterile, 2pcs Packed; REF: 518CS01, Minitip, Ster. Single Wrapped, Molded bp 100mm; REF: 220251, P/N: 518CS01.BD, Minitip, sterile Single Wrapped Molded bp 100mm; REF: 519C/519C.BX, Regular Sterile, Molded break point 100 m

Reason for recall

A sterility assurance level of 10-6 cannot be guaranteed due to intentional fraud and data integrity issues at the contract sterilizer, Steril Milano.

Distribution

US nationwide distribution, including Puerto Rico.

Key facts

Status
Terminated
Initiation date
2021-07-09
Report date
2021-09-15
Termination date
2024-01-25
Voluntary/Mandated
Voluntary: Firm initiated
Location
Brescia, N/A, Italy

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2440-2021