FDA recall Z-2444-2019

Angiodynamics, Inc. · Class II · device

Product

AngioDynamics VenaCure EVLT NeverTouch Procedure Kit, 45CM NEVERTOUCH FRS .018 PROCEDURE KIT PG, Catalog Number 11403012

Reason for recall

The affected product is properly labelled with an expiration date of March 31, 2022; but the RFID programmed expiration date is April-2014 which results in an error message when connected to the laser source and prevents the product from being used.

Distribution

The products were distributed to the following US states: DC, IL, IN, MI, NJ, TN, TX, and WI.

Key facts

Status
Terminated
Initiation date
2019-06-27
Report date
2019-09-11
Termination date
2020-07-29
Voluntary/Mandated
Voluntary: Firm initiated
Location
Queensbury, NY, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2444-2019