FDA recall Z-2444-2021

Roche Diagnostics Operations, Inc. · Class II · device

Product

cobas infinity central lab running software version 3.01.03 through 3.02.08, Catalog#: 07154003001

Reason for recall

Under specific circumstances created by the user, the cobas e flow test results could be replaced by an automatic result sent to the lab LIS. This automatic result could be released and interpreted by your Laboratory Information System (LIS) as a false positive or negative result. It should be noted that the "correct" result would still be available on the instrument.

Distribution

US Nationwide distribution

Key facts

Status
Terminated
Initiation date
2021-06-24
Report date
2021-09-22
Termination date
2023-07-21
Voluntary/Mandated
Voluntary: Firm initiated
Location
Indianapolis, IN, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2444-2021