FDA recall Z-2448-2023

Randox Laboratories Ltd. · Class II · device

Product

Randox Liquid Protein Calibrators-SP CAL-(LIQ)-In vitro diagnostic product used for the calibration of ASO, Complement C3, Complement C4, CRP, Ferritin, Haptoglobin, IgM, Prealbumin, RF and Transferrin assays Catalogue Number: IT2691

Reason for recall

As part of the firm's ongoing quality monitoring (see internal complaint # OCC72574), Randox Laboratories have restandardized Ferritin in Liquid Protein Calibrators, IT2691, to reference material NISBC 19/118. The calibrators lots 2112IT-2116IT, packed into batches 627222, 627224, 634886 and 634887 have been reassigned as part of the restandardization. Following this restandardization, Ferritin results for Quality Control material and patient samples recovered erroneously higher than the targeted calibrator values by approximately +10% across the assay range following.

Distribution

CA, FL, IL, ME, MI, NC, OH, PA,NJ,

Key facts

Status
Ongoing
Initiation date
2023-07-19
Report date
2023-08-30
Voluntary/Mandated
Voluntary: Firm initiated
Location
Crumlin (North), N/A, Ireland