FDA recall Z-2448-2024

ICU Medical Inc · Class I · device

Product

Replacement Battery List Number SUB0000594 and SUB0000864 found in Plum A+ & Plum A+3 Infusion Systems, Pump List Numbers: 11005, 11971, 12391, 12618, 20678, 20679, 20792, 60529, 12348, 11973. OUS units marketed with the following descriptions: Plum A+ Mednet TM Firmware New 13.6 Span, Plum A+3, PLUM A+3 INFUSION PUMPS and Plum A+3 Wireless Refurb, PLUM A+, PLUM A+ Infusion System, PlumA+3IV Pump Multi 1' Set

Reason for recall

Due to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the battery lifecycle than expected.

Distribution

Worldwide Distribution: US (nationwide) and OUS (foreign) countries of: Australia, Canada, France, Guadeloupe, Guatemala, Ireland, Italy, Oman, Paraguay, Qatar, Spain, Saudi Arabia, South Africa, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Uruguay.

Key facts

Status
Ongoing
Initiation date
2024-05-07
Report date
2024-08-21
Voluntary/Mandated
Voluntary: Firm initiated
Location
Lake Forest, IL, United States