FDA recall Z-2451-2023

Siemens Healthcare Diagnostics, Inc. · Class II · device

Product

ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein (UCFP), Material Number 11319151, in vitro diagnostic use in the quantitative determination of total protein in human urine and cerebrospinal fluid

Reason for recall

There is a potential for ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein reagent carryover impacting Enzymatic Creatinine_2 (ECRE_2). Falsely depressed ECRE_2 results may be observed when the assay is processed after the UCFP test on ADVIA Chemistry systems.

Distribution

Worldwide - US Nationwide distribution in the states of CA, CO, CT, FL, GA, IA, MD, MI, NJ, NM, NY, OH, PA, SD, TX, UT, and VA. Global distribution.

Key facts

Status
Ongoing
Initiation date
2023-07-25
Report date
2023-08-30
Voluntary/Mandated
Voluntary: Firm initiated
Location
Tarrytown, NY, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2451-2023