FDA recall Z-2456-2019

Allergan PLC · Class I · device

Product

Natrelle133 Plus Tissue Expander With/Without Suture Tabs, BioCell Full Height Variable Projection with Magna-Fider Xact & 21G Needle Infusion Set, Sterile, for the following style numbers: STYLE 133P-FV: 133P-FV-11, 133P-FV-12, 133P-FV-13, 133P-FV-14, 133P-FV-15, 133P-FV-16; STYLE 133P-MV: 133P-MV-11, 133P-MV-12, 133P-MV-13, 133P-MV-14, 133P-MV-15, 133P-MV-16; STYLE 133P-LV: 133P-LV-11, 133P-LV-12, 133P-LV-13, 133P-LV-14, 133P-LV-15, 133P-LV-16; STYLE 133P-FX: 133P-FX-11, 133P-FX-12, 133P-FX-13, 133P-FX-14, 133P-FX-15, 133P-FX-16; STYLE 133P-MX:133P-MX-11,133P-MX-12, 133P-MX-13, 133P-MX-14, 133P-MX-15, 133P-MX-16; STYLE 133P-SX: 133P-SX-11, 133P-SX-12, 133P-SX-13, 133P-SX-14, 133P-SX-15, 133P-SX-16; STYLE 133P-SV: 133P-SV-11,133P-SV-12, 133P-SV-13, 133P-SV-14, 133P-SV-15, 133P-SV-16; STYLE 133P-FV-T:133P-FV-11-T, 133P-FV-12-T, 133P-FV-13-T, 133P-FV-14-T, 133P-FV-15-T, 133P-FV-16-T; STYLE 133P-MV-T: 133P-MV-11-T, 133P-MV-12-T, 133P-MV-13-T, 133P-MV-1

Reason for recall

The action was initiated following notification by the U.S. FDA of their recently updated, global safety information concerning the higher incidence of anaplastic large cell lymphoma (BIA-ALCL) in patients who have textured breast implants.

Distribution

Worldwide Distribution and US Nationwide All fifty (50) US States, the US Virgin Islands and Puerto Rico. (AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY). OUS: Bermuda, Canada, Chile, Japan, Korea, Thailand, Taiwan and Vietnam.

Key facts

Status
Ongoing
Initiation date
2019-07-24
Report date
2019-09-18
Voluntary/Mandated
Voluntary: Firm initiated
Location
Irvine, CA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2456-2019