FDA recall Z-2458-2025

mo-Vis BVBA · Class I · device

Product

All-round Joystick R-net Light. Electrical wheelchair component.

Reason for recall

Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.

Distribution

US Nationwide distribution in the state of TX.

Key facts

Status
Ongoing
Initiation date
2025-07-10
Report date
2025-09-10
Voluntary/Mandated
Voluntary: Firm initiated
Location
Nevele, N/A, Belgium

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2458-2025