# FDA recall Z-2460-2021

> **CAREFUSION** · Class II · device recall initiated 2021-08-02.

## Product

BD  Extension Set SMALLBORE Tubing SmartSite VALVE, REF: 10010511; BD SmartSite Extension Set Needle-Free Valve, REF: 10010912; BD EXTENSION SET SMALLBORE TUBING DETACHABLE SmartSite VALVE, REF: 10010983; BD Gravity Burette Set 60 DP 4 VersaSafe PORTS 3 SmartSite VALVES BALL VALVE DRIP CHAMBER V/NV, REF: 10011087; BD EXTENSION SET SMALLBORE TUBING SmartSite VALVE (DETACHABLE), REF: 10011253; BD EXTENSION SET LOW SORBING TUBING 0.2 MICRON FILTER SmartSite VALVE, REF: 10013902; BD EXTENSION SET 2 SmartSite VALVES Y-CONNECTOR, REF: 10014692; BD EXTENSION SET SMALLBORE TUBING 2 DETACHABLE SmartSite VALVES, REF: 10015488; BD SmartSite NEEDLE FREE VALVE, REF: 2000E; BD Extension Set J-LOOP SMALLBORE Tubing SmartSite VALVE, REF: 20021E; BD Extension Set SMALLBORE Tubing SmartSite VALVE, REF: 20035E; BD Extension Set SMALLBORE Tubing SmartSite VALVE SLIP LUER, REF: 20037E; BD Extension Set TRI-PORT SMALLBORE Tubing 3 SmartSite VALVES, REF: 20038E; BD Extension Set SMALLBORE Tubing SmartSite VA

## Reason for recall

Needle-free connector valves, either standalone or attached to extension sets, may have issues including difficulty to flush, flow issues, and partial or total occlusions, which could lead to therapy delay. Follow-up activates not required if product has already been used or was safely connected to the female luer of connecting device.

## Distribution

Worldwide distribution - US Nationwide distribution in the states of FL, PA, NC, OH, SC, IL, RI, MO, NY, TN, AL, VA, PR, CA, NE, NJ, MI, MD, NM, LA, KY, NV, ME, OK, WA, MA, WV, TX, AR, OR, KS, GA, VT, IN, WY, AZ, CO, MN, WI, UT, SD, MS, ID, MT, ND, HI, IA, DC, AK, NH, CT, DE, GU and the countries of CA, NZ, TW, AU, BE, VN, AR, MY, TH, KR, ID, PH, SG, BH, SA, HK, IN, CN, BR.

## Key facts

- **Recall number:** Z-2460-2021
- **Recalling firm:** CAREFUSION
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2021-08-02
- **Report date:** 2021-09-22

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** San Diego, CA, United States

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2460-2021