FDA recall Z-2471-2023

Siemens Healthcare Diagnostics, Inc. · Class II · device

Product

Atellica CH Total Bilirubin_2 -In vitro diagnostic use in the quantitative determination of total bilirubin in human serum and plasma (lithium heparin) of adults and neonates Test Code: TBil_2 Siemens Material Number (SMN): 11097531

Reason for recall

Potential for reagent carryover on the Atellica CH 930 resulting in a positive bias that could impact quality control (QC), patient samples, and calibrator results with from Atellica¿ CH Total Bilirubin_2 (TBil_2), Atellica¿CH LDL Cholesterol (LDLC), Atellica¿ CH Gamma-Glutamyl Transferase (GGT) and Atellica¿CH HDL Cholesterol (HDLC)

Distribution

US Nationwide - Worldwide Distribution Foreign: Taiwan Uruguay Vietnam

Key facts

Status
Ongoing
Initiation date
2023-07-10
Report date
2023-09-06
Voluntary/Mandated
Voluntary: Firm initiated
Location
Tarrytown, NY, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2471-2023