FDA recall Z-2475-2019

Alere San Diego, Inc. · Class II · device

Product

Alere Cholestech LDX Analyzer, model # 14-874 containing power supplies with the identifying codes of 1837d and 1843d - Product Usage: The Alere Cholestech LDX System is a small, portable analyzer and test cassette system. The System is for in vitro diagnostic use only. The Lipid Profile GLU Cassette is for the quantitative determination of total cholesterol, HDL (high-density lipoprotein) cholesterol, triglycerides and glucose in whole blood. A TC/HDL (total cholesterol/HDL cholesterol) ratio and estimated values for LDL (low-density lipoprotein) and non-HDL cholesterol are calculated by the Alere Cholestech LDX¿ Analyzer.

Reason for recall

The polarity of the output of power supply was reversed. Due to polarity of the output of power supplies being reversed, it renders the power supplies nonfunctional and the analyzer inoperable.

Distribution

US Nationwide distribution in the states of AK, A, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NY< OH, OR, PA, RI, SC, SD, TN, TX, VA, VT, WA, WI. No OUS consignees

Key facts

Status
Terminated
Initiation date
2019-02-15
Report date
2019-09-11
Termination date
2023-12-12
Voluntary/Mandated
Voluntary: Firm initiated
Location
San Diego, CA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2475-2019