FDA recall Z-2478-2025

WASATCH PHOTONICS · Class II · device

Product

Wasatch Photonics 785/830-SR and 785-ER-ILC Raman Spectrometers, Product Models: 785-SR-ILP, 785-SR-ILC, 830-SR-ILP, 830-SR-ILC,785-ER-ILC.

Reason for recall

Two new potential laser safety failure modes have been identified in the product design. Overcurrent of laser diode, and a fail-unsafe mode. The failure modes have never been found to occur in these models, whether in the field or factory, but could potentially occur under excessive force or extreme, continuous vibration.

Distribution

U.S.

Key facts

Status
Ongoing
Initiation date
2025-08-04
Report date
2025-09-17
Voluntary/Mandated
FDA Mandated
Location
Morrisville, NC, United States

Primary source