FDA recall Z-2479-2020

Preventice Services, LLC · Class II · device

Product

BodyGuardian Heart Remote Monitoring Kit, consisting of two BG Heart monitors, either Models BC-1001-A or BC1001-D, and a cell phone, either Models BP1008 or BP10014, that is paired to both of the monitors.

Reason for recall

The device data being collected and transferred to the monitoring center may not be accurate due to nonvalidated association between the phone software and the heart monitors, therefore, the patient's report should not be used to evaluate their condition.

Distribution

US Nationwide Distribution to AK, AL, GA, NJ, OH, and TX. There was no foreign/military/government distribution.

Key facts

Status
Terminated
Initiation date
2019-08-06
Report date
2020-07-08
Termination date
2022-11-18
Voluntary/Mandated
Voluntary: Firm initiated
Location
Houston, TX, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2479-2020