FDA recall Z-2489-2023

Accuray Incorporated · Class II · device

Product

Upper roller bracket (part number 1088370) for the Gen2 gantry enclosure for the Radixact Treatment Delivery System.

Reason for recall

It is possible the upper roller bracket in the Radixact System gantry enclosure may not provide enough clearance for service personnel to use a jib crane for moving components while servicing the system, resulting in possible injury to the service personnel.

Distribution

Worldwide - US Nationwide distribution in the states of CA, GA, IL, LA, NJ, OR, PA, SC, SD, TX, and WA. There was government distribution and no military distribution. The countries of Australia, Belgium, Canada, Colombia, France, India, Indonesia, Italy, Japan, Kazakhstan, Malaysia, Poland, Saudi Arabia, South Korea, Spain, Switzerland, Taiwan, Thailand, Turkey, and United Kingdom.

Key facts

Status
Completed
Initiation date
2023-08-03
Report date
2023-09-06
Voluntary/Mandated
Voluntary: Firm initiated