# FDA recall Z-2522-2020

> **Shanghai United Imaging Healthcare Co., Ltd.** · Class II · device recall initiated 2020-05-08.

## Product

uCT 760 System, Computed Tomography X-ray System

## Reason for recall

Two  issues were identified with the computed tomography x-ray system including an intermittent issue which may affect image data leading to artifacts; and intermittent scout scanning interruption due to occasional angle signal drift.  This  may result in the need for rescanning of the patient with the patient receiving an additional dose of radiation and possible use of additional contrast media.

## Distribution

Distributed to two customers. Both customers located in Houston, Texas.

## Key facts

- **Recall number:** Z-2522-2020
- **Recalling firm:** Shanghai United Imaging Healthcare Co., Ltd.
- **Classification:** Class II
- **Product type:** device
- **Status:** Terminated
- **Initiation date:** 2020-05-08
- **Report date:** 2020-07-15
- **Termination date:** 2022-01-20

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Shanghai, China

## Primary source

FDA Enforcement Reports: https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2522-2020