FDA recall Z-2525-2018

Siemens Medical Solutions USA, Inc · Class II · device

Product

AXIOM Artis with Laird Cooling Unit for SSFD: Material # 7555118. Artis is an angiography system developed for single and biplane diagnostic imaging and interventional procedures.

Reason for recall

Due to a defective sealing, coolant may inflow into the electrical parts of the system cabinet, which could cause a short circuit and cause the system to shut down suddenly during an ongoing procedure

Distribution

HI , IL, MI, NC, NY, VA, WI

Key facts

Status
Terminated
Initiation date
2018-05-25
Report date
2018-08-01
Termination date
2019-05-16
Voluntary/Mandated
Voluntary: Firm initiated