FDA recall Z-2525-2025

ICU Medical Inc. · Class II · device

Product

Smiths Medical CADD" Medication Cassette Reservoir with flow stop, clamp and female Luer. Non-vented stopper included. 50mL, Product Code 21-7301-24

Reason for recall

Certain CADD 50 mL and 100 mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter.

Distribution

Worldwide distribution.

Key facts

Status
Ongoing
Initiation date
2025-08-06
Report date
2025-09-10
Voluntary/Mandated
Voluntary: Firm initiated
Location
Minneapolis, MN, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2525-2025