FDA recall Z-2532-2020

Bound Tree Medical · Class II · device

Product

Curaplex Ambu King LTS-D Supraglottic Airway Kit with ET Tube Part Number: KLTSD404K Curaplex Ambu¿ King LTS-D" Supraglottic Airway Kit with ET Tube (hereafter the Kit which contains all of the components necessary to successfully obtain an open patient airway with an Ambu¿ King LTS-D" Supraglottic Airway.

Reason for recall

Kit contained an incorrect catheter-tip syringe not compatible with the luer lock connector of the included airway. As a result, may be unable to inflate the balloons of the supraglottic airway to create a secure seal in the patient s airway as intended by the kit

Distribution

Nationwide Distribution

Key facts

Status
Terminated
Initiation date
2020-05-14
Report date
2020-07-15
Termination date
2023-12-07
Voluntary/Mandated
Voluntary: Firm initiated
Location
Dublin, OH, United States