FDA recall Z-2539-2020

OrthoPediatrics Corp · Class II · device

Product

OrthoPediatrics ACL Reconstruction System ShieldLoc 7mm implant component model number 10-1008-4070 - Product Usage: is intended for fixation of tendons and ligaments during orthopedic reconstruction procedures such as Anterior Cruciate Ligament (ACL) Reconstruction.

Reason for recall

ShieldLoc implant component of the ACL Reconstruction Systemlots may be labeled with the incorrect size. Product labeled as 10-1008-4070 was found to have 10-1008-4060 product in the package. Product labeled as 10-1008-4060 was found to have 10-1008-4070 product in the package.

Distribution

US Nationwide distribution.

Key facts

Status
Terminated
Initiation date
2020-06-05
Report date
2020-07-15
Termination date
2021-05-25
Voluntary/Mandated
Voluntary: Firm initiated
Location
Warsaw, IN, United States