# FDA recall Z-2545-2024

> **Beckman Coulter, Inc.** · Class II · device recall initiated 2024-06-14.

## Product

Discrete photometric chemistry analyzer for clinical use.

## Reason for recall

Beckman Coulter, Inc. has identified a rise in reagent carousel motion errors in Access 2 Immunoassay System Analyzers that appear to be caused by misaligned index wheels housed within the reagent carousel module. Customers affected by this issue may observe an increase in reagent carousel motion errors or QSMC errors that occur while unloading or loading reagent packs during system initialization and sample processing, which causes a non-operational system status and cancels tests; and patient results may be delayed if these errors cancel test processing.

## Distribution

Worldwide - US Nationwide distribution in the states of NE, MT, LA, TX, NY, CA, MI, AL, IL the countries of Austria, Belgium, Georgia, India, Indonesia, Italy, Korea, Malaysia, Myanmar, Nigeria, Switzerland, Vietnam.

## Key facts

- **Recall number:** Z-2545-2024
- **Recalling firm:** Beckman Coulter, Inc.
- **Classification:** Class II
- **Product type:** device
- **Status:** Ongoing
- **Initiation date:** 2024-06-14
- **Report date:** 2024-08-21

- **Voluntary/Mandated:** Voluntary: Firm initiated
- **Location:** Chaska, MN, United States