FDA recall Z-2551-2020

Roche Molecular Systems, Inc. · Class II · device

Product

MagNA Pure 96 Instrument, clinical sample concentrator, Material number 06541089001 - Product Usage: This protocol is intended for pipetting samples into a MagNA Pure 96 Processing Cartridge (purification rack position 1) before a purification run.

Reason for recall

When using Sample Transfer protocol version 3.0, the drop catcher is not activated on the MagNA Pure 96 instrument, which introduces a potential risk for cross-contamination leading to erroneous results.

Distribution

Worldwide distribution - US Nationwide distribution.

Key facts

Status
Terminated
Initiation date
2020-06-03
Report date
2020-07-22
Termination date
2023-10-04
Voluntary/Mandated
Voluntary: Firm initiated
Location
Branchburg, NJ, United States

Primary source