FDA recall Z-2556-2018

CryoLife, Inc. · Class II · device

Product

BioGlue Spreader Tip, 12mm, Sterile, RX Only. Product Usage: BioGlue is indicated for use as an adjunct to standard methods of achieving hemostasis (such as sutures and staples) in adult patients in open surgical repair of large vessels (such as aorta, femoral , and carotid arteries).

Reason for recall

The CryoLife L6318 (Label, BioGlue 12mm Spreader Tip, Indicator Label 2 1/8" x 1 1/4") was incorrectly applied to the Syringe Applicator Tip pouches and Syringe Spreader Tip pouches were not included in the boxes.

Distribution

Distribution is to Japan

Key facts

Status
Terminated
Initiation date
2018-06-04
Report date
2018-08-01
Termination date
2019-02-08
Voluntary/Mandated
Voluntary: Firm initiated
Location
Kennesaw, GA, United States

Primary source

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2556-2018